Circulating coolers

Available with cooling capacities of up to 2000W and temperature ranges of -30°C to +40°C:
Our circulating coolers can be used throughout scientific and industrial laboratories for various cooling applications due to their universal properties. The compact, space-saving design and the very quiet cooling unit allows users to place our coolers directly on the desk in the workplace.

Circulating Cooler DLK 402

The smallest in its range. The DLK 402 circulating cooler serves as a universal cooler for the entire laboratory. A new method of casing construction provides for modern design, low weight and guarentees innovative technical features. Due in particular to the compact space-saving design with low-noise cooling unit, the freezer unit is suitable for direct use on the desk in the workplace.

Small size, great for cooling - Circulating cooler DLK 402

Model DLK 402
Operation -10°C to +40°C
Control accuracy +/- 1 K
Cooling capacity in watts
at +30 °C 380
at +10 °C 270
Circulation pump
Output 4l/min
Discharge head 
1,5mWs
External dimensions [WxDxH]
26 x 37 x 40,5 cm
Weight
18,2kg
Ambient temperature +12°C to +30°C
Electrical connection 230V / 50Hz
Max. current
1,5A
Coolant tank 4l
Ask for a quote favicon.ico
  • Electronic temperature control with LED display
  • Float switch for monitoring coolant level
  • Level indicator on the front of the appliance
  • Limit alarm when the permitted working area is exceeded
  • Refigeration unit: low noise, fully hermetically sealed, air-cooled, low maintenance
  • Different hose connections included
  • 3/4“ filler neck and vent valve on the top of the appliance
  • Casing made of rigid polyurethane foam
  • Heat exchanger in non-ferrous metal version

Thermo-insulated hoses

  • Silicon Hoses: ID 10,0mm // to 0.5 bar // to -30°C // without fabric
  • Length on demand (maximum length: 10m)

TP Thermal plate 

For cooling and stabilising temperatures in conjunction with FRYKA DLK circulating coolers

  • two hose fittings made of stainless steel (suitable for DLK)
  • Stainless steel plates
  • Processing of refrigerated products during cooling possible
  • Low height:
    - ergonomic work
    - can be placed on table
    - for installation
  • Seperate installation of refrigeration unit and thermal plate possible
  • Plate size as desired

further information

 

 

HKF coolant 15.1

  • Recommended for all FRYKA circulating coolers
  • 5 litre container
  • Ready-made mixture
  • Frost-proof up to -15°C

DLK 402DP:
dual pump with increased output to 6l/min and discharge head to 3mWs

 
 
 

Circulating cooler to 2000 watts cooling capacity

FRYKA circulating coolers are mainly used by our customers where cooling systems have so far been operated with industrial water. Some examples are the removal of process heat, stabilising of the temperature of laboratory equipment such as centrifuges, analysers, rotary evaporators or reaction vessels. All DLK circulating coolers can be used as closed or open systems.

The right appliance for every need - FRYKA DLK circulating coolers

Modell DLK 602 DLK 632 
DLK 1002
DLK 2002
Operation -10°C to +40°C -30°C to +20°C -10°C to +40°C -10°C to +40°C
Control accurancy +/- 1 K
Cooling capacity in watts
at +20 °C 550 550 940 2000
at +10 °C 480 440 780 1500
at -10 °C 180 280 320 600
Circulation pump
Output 12l/min 12l/min 12l/min 12l/min
Discharge head 
6mWs 6mWs 26mWs
30mWs
External dimensions [WxDxH] [cm] 31 x 36 x 52 31 x 36 x 52 34 x 40 x 60 44 x 47 x 76
Weight
32 kg
32 kg
41 kg
80 kg
Electrical connection 230V / 50Hz
Max. current 
2,8A 2,8A 3,5A 8,0A
Coolant tank 2,6l 2,6l 4,6l 9l
Ask for a quote favicon.ico favicon.ico favicon.ico favicon.ico
  • Electronic temperature control with LED display
  • Float switch for monitoring coolant level
  • Limit alarm when the permitted working area is exceeded
  • Refrigeration unit: low noise, fully hermetically sealed, air-cooled, low maintenance
  • In DLK 2002 the cooling unit consists of two seperate sets of refrigeration systems. This results in better adjustment and the possibility of emergency operation of the second machine if one compressor breaks down.
  • 12mm hose connections with internal thread (M10x1)
  • 3/4“ filler neck and vent valve on the top of the appliance
  • Casing parts made of stainless steel
  • Heat exchanger in non-ferrous metal version
  • Ram protection ring going over and under

Thermo-insulated hoses

  • Silicon Hoses: ID 12.5mm // to 6.0 bar // to -30°C // with fabric
  • Length on demand (maximum length: 10 m)

TP thermal plate

for cooling and stabilising temperatures in conjunction with FRYKA DLK circulating coolers

  • Two hose fittings made of stainless steel (suitable for DLK)
  • Stainless steel plates
  • Processing of refrigerated products during cooling possible
  • Low height:
    - ergonomic work
    - can be placed on table
    - for installation
  • Seperate installation of refrigeration unit and thermal plate possible
  • Plate size as desired 

further information

  

 

HKF coolant 15.1

  • Recommended for all FRYKA circulating coolers
  • 5 litre container
  • Ready-made mixture
  • Frost-proof up to -15°C
  • Connection for external temperature sensor (with switch intern-extern, incl. sensor PT100 with 2m silicon cable)
  • 4 fixed rollers, 2 with brake
  • Your desired option is not listed? Just ask us!